Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Thermal Vacuum Outgassing Analysis of Materials: TML and CVCM Testing

Thermal vacuum outgassing analysis measures the Total Mass Loss (TML) and Collected Volatile Condensable Materials (CVCM) of engineering materials under simulated space and high-vacuum environments. Performed in accordance with ASTM E595, ECSS-Q-ST-70-02 and related ISO test protocols, this analysis is a mandatory qualification step for spacecraft materials and a critical contamination-control tool for optical instruments and vacuum equipment. Accredited third-party laboratories provide high-precision microgravimetric testing to verify that materials meet the widely accepted acceptance criteria of TML below 1.0% and CVCM below 0.10%.

Technical Overview

Outgassing is the release of adsorbed moisture, residual solvents, plasticizers, low-molecular-weight additives, and decomposition products from a material when it is exposed to vacuum and elevated temperature. In the vacuum of space, these liberated molecules travel in straight lines and readily recondense on colder surfaces, forming molecular contamination films that degrade the performance of optical components, thermal control coatings, solar cells, sensors, and cryogenic instruments. Thermal vacuum outgassing analysis quantifies this behavior through accelerated screening under standardized conditions, providing the data needed to select and qualify materials for spacecraft, satellites, and high-vacuum industrial systems.

The industry-standard screening method is ASTM E595, which has been adopted with minor variations by the European standard ECSS-Q-ST-70-02 and referenced in numerous NASA and ISO material qualification programs. The test exposes preconditioned specimens to a vacuum of 5 × 10-5 torr (approximately 6.7 × 10-3 Pa) or less at 125 °C for 24 hours, while a chromium-plated collector plate maintained at 25 °C captures the volatile condensable fraction. Precision microbalance measurements before and after exposure yield two key parameters:

  • TML (Total Mass Loss) – the percentage of specimen mass lost during the vacuum exposure, indicating the overall volatility of the material.
  • CVCM (Collected Volatile Condensable Materials) – the percentage of specimen mass that recondensed on the 25 °C collector, indicating the contamination risk to nearby sensitive surfaces.

For most space applications, materials are accepted when TML is below 1.0% and CVCM is below 0.10%. Materials exceeding these limits require waivers, protective design measures, or replacement with low-outgassing alternatives.

Testing Samples

Polymeric and Elastomeric Materials

Engineering plastics, rubbers, O-rings, gaskets, cable jackets, and flexible membranes are the most outgassing-sensitive material class. Additives such as plasticizers, antioxidants, and curing residues are the dominant sources of volatile species, making these polymers priority candidates for screening.

Adhesives, Sealants and Potting Compounds

Structural adhesives, silicone and epoxy sealants, conformal coatings, and encapsulation or potting compounds used in electronics assemblies are routinely tested. Incomplete cure, residual solvent, and low-molecular-weight siloxanes can produce significant CVCM values, particularly for silicones.

Coatings, Paints and Thin Films

Thermal control paints, black anodizing sealers, optical coatings, metallized polymer films, and multilayer insulation (MLI) materials are tested because condensable contamination directly alters their absorptance, emittance, and reflectance properties.

Tapes, Films and Insulation Products

Pressure-sensitive tapes, polyimide and PTFE films, cable insulation, and blanket materials are common test items, since large exposed surface areas amplify outgassing contributions at spacecraft level.

Lubricants, Compounds and Process Materials

Vacuum greases, dry-film lubricants, locking compounds, and cleaning or molding process residues are screened to ensure they do not become contamination sources during mission life.

Metals, Ceramics and Composites

Fully cured composites, ceramics, and metals generally exhibit very low outgassing; they are tested primarily when they incorporate organic binders, impregnants, surface treatments, or bonded interfaces that may release volatiles.

Testing Items

A comprehensive thermal vacuum outgassing analysis program typically includes the following measured parameters:

  • Total Mass Loss (TML) at 125 °C / 24 h / vacuum, expressed as percent of initial specimen mass.
  • Collected Volatile Condensable Materials (CVCM) on a 25 °C collector plate, expressed as percent of initial specimen mass.
  • Water Vapor Regained (WVR) – mass regain after 24 hours of reconditioning at 23 °C and 50% relative humidity, used to separate absorbed moisture from permanent mass loss.
  • Material residual mass after moisture correction, reported as TML minus WVR for moisture-dominated samples.
  • Extended outgassing kinetics (optional) using quartz crystal microbalances (QCM) per ASTM E1559, characterizing condensable flux as a function of time and temperature for contamination modeling.
  • Re-test and verification of materials after thermal cycling, humidity aging, or post-curing treatments to confirm stability.

Testing Methods

The reference screening procedure is performed according to ASTM E595 and ECSS-Q-ST-70-02, with the following workflow:

  • Specimen preparation: typically three specimens per material, each 100–300 mg, cut or prepared to expose representative surfaces.
  • Preconditioning: 24 hours at 23 ± 2 °C and 50 ± 5% relative humidity, followed by gravimetric measurement on a microbalance with resolution at the microgram level.
  • Vacuum exposure: specimens are placed in a heated chamber at 125 °C for 24 hours under a vacuum of 5 × 10-5 torr or lower; effusing molecules impinge on a collector maintained at 25 °C positioned above each specimen.
  • Collector weighing: the collector plates are weighed to determine the mass of recondensed volatiles, yielding the CVCM value.
  • Reconditioning and WVR: specimens are stabilized for 24 hours at standard laboratory conditions, reweighed to determine water vapor regained, and the final TML is calculated.
  • Reporting: results are compared against acceptance criteria (TML ≤ 1.0%, CVCM ≤ 0.10%), with deviations, specimen descriptions, and measurement uncertainties documented.

Where mission requirements demand deeper insight, supplementary methods are applied: ASTM E1559 QCM-based testing at multiple temperatures provides time-dependent outgassing and recondensation data for contamination transport modeling, and thermogravimetric analysis (TGA) coupled with mass spectrometry can identify the chemical nature of the evolved species. Alternative exposure temperatures (for example 150 °C or 200 °C) may be selected for materials intended for higher-temperature service, provided the test conditions are clearly documented.

Applications

Thermal vacuum outgassing analysis supports material qualification and contamination control across a wide range of high-technology sectors:

  • Spacecraft and satellite material selection, including qualification of all polymeric materials flying on launch vehicles, satellites, and deep-space probes.
  • Optical payload assurance – telescopes, star trackers, imaging spectrometers, and laser systems where molecular films would reduce throughput and increase scatter.
  • Thermal control subsystems – verification of paints, coatings, and multilayer insulation that must retain their optical properties throughout mission life.
  • Electronics and harness assemblies – screening of conformal coatings, adhesives, connectors, and wire insulation used in vacuum environments.
  • Semiconductor and vacuum equipment industries – material screening for deposition chambers, vacuum furnaces, cryopumps, and UHV research systems.
  • Research and development – evaluation of new low-outgassing formulations, post-curing processes, and surface treatment effectiveness.

FAQ

What is thermal vacuum outgassing and why is it important?

Outgassing is the release of volatile molecules from a material under vacuum. In space there is no atmosphere to flush these molecules away, so they recondense on cold surfaces such as optics and detectors, forming contamination films that degrade performance. Screening materials in advance prevents costly mission-level contamination problems.

What are the standard acceptance criteria?

The universally applied criteria from ASTM E595 are TML less than 1.0% and CVCM less than 0.10%. Some programs with sensitive optics or cryogenic instruments impose tighter limits, for example CVCM below 0.05%, on materials located near critical surfaces.

What are the standard test conditions?

The reference condition is 125 °C for 24 hours at a pressure of 5 × 10-5 torr or lower, with a condensable collector at 25 °C. Specimens are preconditioned for 24 hours at 23 °C and 50% relative humidity before exposure.

How much sample is required?

Typically 100–300 mg per specimen, with three specimens per material. Larger or assembled components can be tested by preparing representative coupons cut from the item, ensuring all material layers and bonded interfaces are included.

Can a material that fails be made acceptable?

In many cases yes. Extended vacuum baking (post-curing or degassing) at elevated temperature can reduce TML and CVCM significantly, especially for adhesives and elastomers. Retesting after such treatment verifies whether the material now meets the acceptance criteria.

How long does the analysis take?

A standard ASTM E595 test, including preconditioning, vacuum exposure, reconditioning, and weighing, is generally completed within one week. Additional conditioning treatments, extended QCM studies, or repeat verification will lengthen the schedule accordingly.

Which standards govern this testing?

The primary standards are ASTM E595, ECSS-Q-ST-70-02, and ASTM E1559 for QCM-based kinetics, with related requirements referenced in ISO spacecraft material qualification documents and national aerospace material handbooks.